3-37
Graphical User Interface
5. To confirm the changes, tap the
OK
button in the pop-up window that appears. Tapping the
X
icon closes
this window and exits without saving.
You will be returned to the “Controls” screen.
If you have chosen “Advanced Mode”, a click-wheel with “Timed Run” options will now be added to the
screen that opens when you tap the
Time
box on the “Home” screen.
If you have reverted to “Standard Mode”, there is no click-wheel with “Timed Run” options.
6. To further customize “Standard” or “Advanced Mode”, please refer to the previous section
View Time As
The
View Time As
button lets you choose whether you wish to have the counter in the time box of the “Home
“screen run up or down while the centrifuge is spinning. The options are:
Elapsed: Makes the timer run forward in continuous mode or up to the time setpoint in a timed run.
Remaining: Makes the timer count down to zero from the time setpoint.
Proceed as follows to toggle between count-up and count-down timer modes:
1. Tap the
Settings
icon on the navigation bar.
2. Tap the
Controls
button on the “Settings” screen.
The “Controls” screen appears.
3. Tap the
View Time As
menu and choose
Elapsed
to have the timer run forwards (factory default), or
Remaining
to have it run backwards.
4. Tap the
Save
button to save the new timer mode.
A pop-up window appears, indicating that the timer mode has been successfully changed.
5. To confirm the changes, tap the
OK
button in the pop-up window that appears. Tapping the
X
icon closes
this window and exits without saving.
You will be returned to the “Controls” screen.
Pulse Customization
The
Pulse Customization
button opens the “Pulse Customization” screen that lets you set the default behavior
for the Pulse button on the “Home” screen. The following options are available for customizing the Pulse
button:
Continuous (factory default): This option makes the centrifuge run continuously at the default speed (for
instructions on presetting the default speed, please refer to the section
) until
you stop it by tapping the Pulse button or Stop button on the “Home” screen.
Max speed: This option makes the centrifuge spin up to maximum speed and then stop.
15sec, 30sec, 1min: These options delay the start of the time counter until maximum speed has been
reached.
Proceed as follows to select the default behavior for the Pulse button :
1. Tap the
Settings
icon on the navigation bar.
2. Tap the
Controls
button on the “Settings” screen.
The “Controls” screen appears.
3. Tap the
Pulse Customization
button.
The “Pulse Customization” screen appears.
4. Choose an option.
5. Tap the
Save
button to save the new mode for the Pulse button .
A pop-up window appears, indicating that the Pulse button has been successfully customized.
6. To confirm the changes, tap the
OK
button in the pop-up window that appears. Tapping the
X
icon closes
this window and exits without saving.
You will be returned to the “Controls” screen.
Lid Auto Open
This menu lets you preset the lid of the centrifuge to remain closed or unlock automatically once the run is
complete. The drop-down list has two options only:
Yes: Lid will unlock automatically once the centrifugation run is complete.
No: Lid will remain closed after the end of the centrifugation run.
Содержание Sorvall ST Plus Series
Страница 1: ...Thermo Scientific Sorvall X Pro ST Plus Series Centrifuges Instructions for Use 50158527 d 08 2019 ...
Страница 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Страница 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Страница 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Страница 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Страница 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Страница 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Страница 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Страница 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Страница 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Страница 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Страница 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Страница 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...