5-4
Maintenance and Care
5. 4. Disinfection
You are responsible that the level of disinfection is achieved according to your requirements.
After disinfection:
1. Rinse the centrifuge and all affected accessories with water.
2. Allow to fully drain and dry.
3. After disinfecting, treat the entire surface of aluminum parts including the cavities with corrosion protection
oil (70009824).
Treat the bolts of swing out rotors with bolt grease (75003786) if necessary.
WARNING
Do not touch infected parts. Hazardous infection is possible when touching
the contaminated rotor and centrifuge parts. Infectious material can get
into the centrifuge when a tube breaks or as a result of spills. In case of
contamination, make sure that no one is put at risk. Disinfect the affected
parts immediately.
CAUTION
Equipment can be damaged by inappropriate disinfection methods
or agents. Make sure that the disinfection agent or the method will
not damage the equipment. In doubt contact the manufacturer of
the disinfection agent. Observe the safety precautions and handling
instructions for the disinfection agents used.
5. 5. Decontamination
You are responsible that the level of decontamination is achieved according to your requirements.
After decontamination:
1. Rinse the centrifuge and all affected accessories with water.
2. Allow to fully drain and dry.
3. After decontaminating, treat the entire surface of aluminum parts including the cavities with corrosion
protection oil (70009824).
Treat the bolts of swing out rotors with bolt grease (75003786) if necessary.
WARNING
Do not touch contaminated parts. Exposure to radiation is possible when
touching the contaminated rotor and centrifuge parts. Contaminated
material can get into the centrifuge when a tube breaks or as a result
of spills. In case of contamination, make sure that no one is put at risk.
Decontaminate the affected parts immediately.
CAUTION
Equipment can be damaged by inappropriate decontamination methods
or agents. Make sure that the decontamination agent or the method will
not damage the equipment. In doubt contact the manufacturer of the
decontamination agent. Observe the safety precautions and handling
instructions for the decontamination agents used.
Содержание Sorvall ST Plus Series
Страница 1: ...Thermo Scientific Sorvall X Pro ST Plus Series Centrifuges Instructions for Use 50158527 d 08 2019 ...
Страница 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Страница 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Страница 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Страница 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Страница 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Страница 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Страница 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Страница 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...
Страница 149: ...B 42 MicroClick 30 x 2 B 13 4 Biocontainment Certificate ...
Страница 152: ...B 45 MicroClick 18 x 5 B 14 4 Biocontainment Certificate ...
Страница 155: ...B 48 Fiberlite F21 48 x 2 B 15 4 Biocontainment Certificate ...
Страница 159: ...B 52 Fiberlite F10 6 x 100 LEX B 17 4 Biocontainment Certificate ...