If the Shandon ClearVue is to be taken out of operation for a long
period of time, or put into storage, ensure the following procedure
has been followed:
Empty the
Mountant Bottle
and clean with
xylene.
Flush the system with xylene (see
Section 2-10
).
Cap the
Mountant Dispense Needle
.
Empty the
Dispense Head Cleaning Station
.
Empty the
Xylene Tray
.
Ensure the instrument has been thoroughly
cleaned and decontaminated as necessary.
If the Shandon ClearVue has been used with, or has come into
contact with, hazardous material, ensure that the appropriate
decontamination procedures have been followed (See World
Health Organization ‘Laboratory Biosafety Manual’).
If the instrument is to be put into storage, re-pack
it into its original packing (see
Appendix C
).
•
•
•
•
•
•
•
5-3 - Taking Out of
Operation and Storage
Содержание Shandon ClearVue A79210100
Страница 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Страница 106: ...106 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT To clean the Slip Wiper 5 2 5 Cleaning the Slip Wiper...
Страница 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Страница 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Страница 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Страница 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Страница 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Страница 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Страница 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...