86
A79210100 - Iss 2
Shandon ClearVue™ Operator Guide
DRAFT
Problem
Cause
Remedy
breaking Slides.
Incorrect size
of slide.
Refer to
Section
1-2
for maximum
permissible
slide sizes.
Chipped or broken
slide loaded
into basket.
Ensure only
undamaged slides
are loaded into
instrument.
Slide loaded into
basket incorrectly.
Ensure slides are
correctly loading
into baskets.
basket slider not
fully closed.
Ensure slider is
fully closed prior
to loading basket.
Damaged or
broken basket.
Replace basket.
build up of
dried mountant
on basket.
Clean or replace
basket.
Dirty camera lens.
Clean camera lens
(
see Section 5-2-6
).
Incorrectly fitted
Transfer head
Suction Cup.
Ensure Suction Cup
is fitted correctly
(
see Section 5-2-4
).
Transfer head
Pads clogged
with mountant.
Clean Transfer
head (
see
Section 5-1
).
Содержание Shandon ClearVue A79210100
Страница 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Страница 106: ...106 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT To clean the Slip Wiper 5 2 5 Cleaning the Slip Wiper...
Страница 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Страница 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Страница 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Страница 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Страница 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Страница 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Страница 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...