90
A79210100 - Iss 2
Shandon ClearVue™ Operator Guide
DRAFT
The following sections give details on how to carry out basic
cleaning and maintenance procedures.
If a problem occurs which is not covered in these sections contact
Thermo Fisher Scientific Service Department.
Rubber
O-Rings
are used to ensure air-tight or water-tight seals
either for the reliable operation of the instrument, or for the safety
of the user.
If these seals become encrusted with mountant their ability to
create a good seal is impaired and they must be replaced.
There are two user serviceable seals on the Shandon ClearVue.
These are located on:
The
Mountant Bottle
spout.
The
Xylene Tray
.
To remove the
O-Ring Seal
on the
Mountant Bottle
spout for
replacement:
Ensure that there are no
Baskets
currently
being processed within the Shandon ClearVue.
Unscrew and remove the lid of the
Mountant
Bottle
.
•
•
1.
2.
5-2 - User Cleaning
and Maintenance
5-2-1 - Changing
Seals
O-Ring
Seal
O-Ring
Seal
Содержание Shandon ClearVue A79210100
Страница 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Страница 106: ...106 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT To clean the Slip Wiper 5 2 5 Cleaning the Slip Wiper...
Страница 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Страница 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Страница 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Страница 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Страница 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Страница 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Страница 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...