118
A79210100 - Iss 2
Shandon ClearVue™ Operator Guide
DRAFT
When the
Flush
cycle is complete, check that
the
Mountant Bottle Cap
is clear of mountant.
Place the
Dispense Head Cap
over the
Dispense Head Needle
to prevent the xylene
which remains in the system from leaking out
during transit.
Approximately 5ml of xylene will remain in the system - follow
good laboratory practice when packing and unpacking.
Disconnect the
Power Lead
.
Ensure the
Battery Switch
has been set to the
off
‘O’
position.
Follow the
Packing Instructions
on the
following page.
•
•
•
•
•