Shandon ClearVue™ Operator Guide
A79210100 - Iss 2
111
DRAFT
Do not disconnect the tubing!
Place the
Mountant Dispense Needle
Assembly
in a container of xylene and allow to
soak overnight.
When the
Needle
is clean, replace the
Mountant Dispense Needle Assembly
to the
Support Arm.
Screw the
Mountant Dispense Needle
Assembly
back in as shown.
Take care not to screw the Mountant Dispense Needle Assembly
in cross-threaded as this will damage the support arm.
The Mountant Dispense Needle Assembly should only be
tightened by hand - do not use a wrench.
•
•
•