84
A79210100 - Iss 2
Shandon ClearVue™ Operator Guide
DRAFT
Problem
Cause
Remedy
breaking
coverslips.
Incorrect size
of coverslip.
Refer to
Section
1-2
for permissible
coverslip options.
broken slip
in hopper.
Clear broken slips
from hopper.
Debris on Slip
Dispense Carriage.
Clean carriage
(see
Section 5-2-7
).
Mountant dried
on Slip Dispense
Carriage.
Clean carriage
(see
Section 5-2-7
).
If this is ineffective
call Thermo Fisher
Service Team
for support.
Loss of vacuum
on Slip Dispense
Carriage.
Call Thermo Fisher
Service Team
for support.
Faulty Coverslip
Transfer head
Suction Cup.
Replace Suction
Cup (see
Section 5-2-4
).
Other error.
Call Thermo Fisher
Service Team
for support.
Содержание Shandon ClearVue A79210100
Страница 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Страница 106: ...106 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT To clean the Slip Wiper 5 2 5 Cleaning the Slip Wiper...
Страница 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Страница 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Страница 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Страница 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Страница 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Страница 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Страница 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...