Problem
Cause
Solution
Mountant
Dispense Needle
dirty / blocked.
Low xylene level
in Dispense head
Cleaning Station.
Refill the Dispense
head Cleaning
Station fully (
see
Section 2-8
).
Quick drying
Mountant has
formed a skin
on the Needle.
Contact your
Thermo Fisher
Product Specialist
for advice.
Dispense head
Cleaning Station
not functioning
correctly.
Call Thermo Fisher
Service Team
for support.