Shandon ClearVue™ Operator Guide
A79210100 - Iss 2
79
DRAFT
Problem
Cause
Solution
The Shandon
ClearVue does
not respond when
the mains power
is switched on.
The Shandon
ClearVue is still
starting-up.
Wait approximately
30 seconds for
the start-up
sequence to finish.
No power supply.
Connect the power
lead and switch on
the mains power
at the socket and
the instrument.
The mains fuses
have blown.
Replace the
mains fuse.
Replace the
instrument fuses.
Note: Only a technically
competent person
should replace fuses.
Other error.
Call Thermo Fisher
Service Team
for support.
The Shandon
ClearVue is
switched ON but
the screen is blank.
Instrument in
Standby mode,
Screensaver is on.
Touch the screen to
exit Standby mode.
Other error.
Call Thermo Fisher
Service Team
for support.
4-3 - Troubleshooting
Tables
Содержание Shandon ClearVue A79210100
Страница 96: ...96 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT 5 2 2 Mountant Jar...
Страница 106: ...106 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT To clean the Slip Wiper 5 2 5 Cleaning the Slip Wiper...
Страница 119: ...Shandon ClearVue Operator Guide A79210100 Iss 2 119 DRAFT C2 Packing Instructions...
Страница 120: ...120 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Appendix D Screen Maps A B...
Страница 121: ...Shandon ClearVue Operator Guide A79210100 Iss 2 121 DRAFT A 1 3 2 4 5 1 2 4 3 5 A...
Страница 122: ...122 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT B 1 3 2 4 5 7 6 8 1 3 5 7 2 4 6 8...
Страница 123: ...Shandon ClearVue Operator Guide A79210100 Iss 2 123 DRAFT Appendix E Declaration of Conformity...
Страница 124: ...124 A79210100 Iss 2 Shandon ClearVue Operator Guide DRAFT Index...
Страница 125: ...Shandon ClearVue Operator Guide A79210100 Iss 2 125 DRAFT...