Thermo Scientific Russell RL060C Conductivity Meter User Guide
10
Chapter IV
Calibration
Important Information on Meter Calibration
Your .meter .has .five .measuring .ranges . .You .can .calibrate .one .point .in .each .of .the .
measuring .ranges .(up .to .five .points) . .If .you .are .measuring .values .in .more .than .one .
range, .make .sure .to .calibrate .each .of .the .ranges .you .are .measuring . .
The .following .table .lists .the .corresponding .conductivity .ranges . .You .should .calibrate .
each .range .using .a .solution .that .falls .between .the .values .in .the .Recommended .
Calibration .Solution .Range .column
Conductivity Range
Recommended Calibration Solution Range
0 .00 .to .20 .00 .μS
6 .00 .to .17 .00 .μS
0 .0 .to .200 .0 .μS
60 .0 .to .170 .0 .μS
0 .to .2000 .μS
600 .to .1700 .μS
0 .00 .to .20 .00 .mS
6 .00 .to .17 .00 .mS
0 .0 .to .200 .0 .mS
60 .0 .to .170 .0 .mS
When .you .recalibrate .your .meter, .old .calibrations .are .replaced .on .a .range .basis . .For .
example, .if .you .previously .calibrated .your .conductivity .meter .at .1413 .μS .in .the .0 .to .
2000 .μS .range .and .you .recalibrate .at .1500 .μS .(also .in .the .0 .to .2000 .μS .range), .the .
meter .will .replace .the .old .calibration .data .(1413 .μS) .in .that .range . .The .meter .will .retain .
all .calibration .data .in .other .ranges . .
To .completely .recalibrate .your .meter, .or .when .you .use .a .replacement .probe, .it .is .best .
to .clear .all .calibration .data . .To .erase .all .the .old .conductivity .calibration .data .completely, .
see .the .
Restore Factory Default Values
section . .