Thermo Scientific Russell RL060C Conductivity Meter User Guide
Chapter VIII
Troubleshooting Guide
Problem
Cause
Solution
Power .on .but . .
no .display
a . . Batteries .not .in .place .
a . . Check .that .batteries .are .
in .place .and .making .good .
contact .
b . Batteries .not .in .correct .
polarity .(+ .and .– .position) .
b . Re-insert .batteries .with .
correct .polarity .
c . Weak .batteries .
c . Replace .batteries .
Unstable .
readings
a . Air .bubbles .in .probe .
a . Tap .probe .to .remove .
bubbles .
b . Dirty .probe . .
b . Clean .the .probe .and .
recalibrate .
c . Probe .not .deep .enough .in .
sample .
c . Make .sure .sample .entirely .
covers .the .probe .sensors .
d . External .noise .pickup .or .
induction .caused .by .nearby .
electric .motor .
d . Move .or .switch .off .
interfering .motor . .
e . Broken .probe .
e . Replace .probe .
Slow .response
a . Dirty ./ .Oily .probe .
a . Clean .probe . .See .
Probe
Care & Maintenance
.