Thermo Scientific Russell RL060C Conductivity Meter User Guide
Automatic Calibration
The .automatic .calibration .allows .you .to .quickly .calibrate .the .meter .to .any .of .the .four .
widely .used .conductivity .calibration .standards . .For .a .list .of .calibration .standards .refer .
to .
Table 1
, .
Selection of Automatic or Manual Calibration
. .
In .the .manual .calibration .mode, .you .can .use .your .own .customized .conductivity .
calibration .standard .to .calibrate .this .meter . .
1 . . .Enter .the .advanced .setup .as .described .in .the .
Advanced Setup Overview
section . .
2 . . .Press .
.or .
.until .
ACAL
.appears .on .the .LCD . .
Press .
. .
3 . . Press .
.or .
.to .select .either .
YES
.or .
no
. .
4 . . .Press .
.to .select . .The .meter .will .take .you .back .to .
the .menu, .
ACAL
. .
5 . . .Press .
.or .
.to .move .to .the .next .menu .or .press . . .
.to .exit .to .measurement .mode . .