6
Conductivity Electrode Information
The .Russell .RL060C .hand-held .meter .is .supplied .with .a .conductivity .electrode .with .
a .BNC .connector . .This .conductivity .electrode .(Cat . .No . .014005) .comes .with .stainless .
steel .rings, .cell .constant .of .K .= .1 .0, .and .a .built-in .temperature .sensor .for .automatic .
temperature .compensation .(ATC) . .Its .specially .designed .Ultem-body .housing .has .good .
chemical .resistant .properties . .It .provides .fast .temperature .response .and .reduces .air .
entrapment, .which .makes .it .easy .to .obtain .accurate, .stable .readings .
The .probe .materials .used .which .have .good .chemical .durability .include:
1 . . Polyetherimide .(Ultem) .– .protective .probe .guard
2 . . Polybutylterphalate .(Valox) .– .sensor .housing
3 . . Stainless .Steel .(SS .304) .– .2 .steel .bands
Proper .use .of .the .probe .is .essential .to .ensure .that .the .optimum .measurement .is .taken .
in .a .short .time . .
The .removable .protective .plastic .probe .guard .is .meant .for .simple .periodic .
maintenance .and .it .must .be .kept .intact .during .measurement .and .calibration .
Always .immerse .the .probe .beyond .the .upper .steel .band .
Note:
• DO NOT remove the protective probe guard during
measurement and calibration as it may affect your
readings.
• We recommend that you do not submerge
the probe above the protective guard. You can
submerge the cable for brief periods of time, but
not continuously.
See the
Probe Care and Maintenance
section for
more information.
Figure 6: Conductivity Electrode