Thermo Scientific Russell RL060C Conductivity Meter User Guide

1

Without ATC (Manual Temperature Compensation)

For .manual .temperature .compensation, .simply .unplug .the .probe’s .phono .jack .(not .
BNC) .from .the .meter . .

To .use .manual .temperature .compensation, .you .need .to .enter .the .temperature .value .
of .your .process .into .the .meter . .This .is .the .value .at .which .the .reading .will .manually .
temperature .compensates . .You .can .select .any .temperature .between .0 .and .50 .°C . .
(32 .to .122 .°F) . .Default .value .is .25 .°C . .

1 . .  .Make .sure .that .the .phono .jack .(for .temperature .

measurement) .is .disconnected .from .the .meter . . .
See .

Figure 7

 . .

2 . .  .Switch .on .the .meter .and .if .necessary, .press .

 .

to .select .temperature .measurement .mode . .See .
the .

Change Conductivity to Temperature 

Measurement Mode 

section . .

3 . .  .Press .

 .to .start .temperature .calibration .process . .

4 . .  .The .“

CA

” .will .appear .momentarily .and .a .temperature .

value .will .start .flashing . .

5 . .  .Check .the .temperature .of .your .sample .using .an .accurate .

thermometer . .Wait .for .the .value .to .stabilize .and .press .

 .or .

 .and .adjust .the .value .to .the .reference .

thermometer .used . .

6 . .  .Press

 

 . .The .[

CO

] .indicator .will .appear .for .1 .5 .seconds, .

and .the .reading .will .stop .flashing . .The .temperature .
calibration .is .successfully .performed . .The .meter .returns .
to .measurement .mode . .

C

°

C

°

C

°

°C

Содержание Russell RL060C

Страница 1: ...Russell RL060C User Guide...

Страница 2: ...riode Quatrode QuiKcheK rf link ROSS ROSS Resolution SAOB SMART AVERAGING Smart CheK SMART STABILITY Stacked Stat Face The Enhanced Lab ThermaSense Triode TRIUMpH Unbreakable pH Universal Access are t...

Страница 3: ...ng Started 2 Chapter III Preparation 4 Chapter IV Calibration 10 Chapter V Measurement 19 Chapter VI Advance Setup Functions 22 Chapter VII Probe and Maintenance 30 Chapter VIII Troubleshooting Guide...

Страница 4: ...34 Chapter XII Conductivity Theory 35 Chapter XIII Assistance and Warranty 38 Chapter XIV Declaration of Conformity 39 Chapter XV WEEE Compliance 40 Addendum 1 Calibration Tips 41 Addendum 2 Calculat...

Страница 5: ...ivity S mS and temperature C This sturdy meter measures up to 5 different ranges with auto ranging capability that switches to appropriate measuring range automatically Your meter includes a conductiv...

Страница 6: ...d C temperature Other annunciators include HO when the HOLD function is activated and LO low battery condition See Figure 2 below LCD and Customized Annunciators for RL060C meter 1 Primary display 2 m...

Страница 7: ...irm values in calibration mode and to confirm selections in setup mode In calibration mode Press to scroll through calibration values In setup mode Press to scroll through the setup sub group programs...

Страница 8: ...m edges of the meter out until it is completely separated from the armor Ensure that the cables of conductivity electrode or temperature probe are not connected 2 To insert meter into rubber armor sli...

Страница 9: ...the direction of the arrow and lift up the cover Note the polarity of the battery before inserting it into position Place the cover back and press down until it locks tight Battery Replacement A LO a...

Страница 10: ...hemical durability include 1 Polyetherimide Ultem protective probe guard 2 Polybutylterphalate Valox sensor housing 3 Stainless Steel SS 304 2 steel bands Proper use of the probe is essential to ensur...

Страница 11: ...on until it unlocks and slide the connector off the socket 3 Insert the mini phono jack of the temperature sensor into the socket on the meter as shown below 4 Unplug the phono jack when not in use or...

Страница 12: ...up Functions section 3 Third screen shows t 25 0 C which is the normalization temperature You can set the normalization temperature to either 25 C or 20 C Refer to the Advance Setup Functions section...

Страница 13: ...User Guide Change Conductivity to Temperature Measurement Mode To switch between conductivity measurement mode and temperature measurement mode simply press The customized annunciator helps indicate...

Страница 14: ...ended Calibration Solution Range 0 00 to 20 00 S 6 00 to 17 00 S 0 0 to 200 0 S 60 0 to 170 0 S 0 to 2000 S 600 to 1700 S 0 00 to 20 00 mS 6 00 to 17 00 mS 0 0 to 200 0 mS 60 0 to 170 0 mS When you re...

Страница 15: ...store it dry calibrate the meter once a month If you take measurements at extreme temperatures calibrate at least once a week Ensure that you use new conductivity standard solutions or sachets during...

Страница 16: ...ibration are Table 1 Conductivity Calibration Standards for Automatic Calibrations Meter Normalization Temperature Calibration Standards Range RL060C 25 C 1 84 S for 0 200 S cm 2 1413 S for 0 2000 S c...

Страница 17: ...he calibration standards listed in Table 1 3 Dip the probe into the calibration standard Immerse the probe tip beyond the upper steel band see Figure 6 Stir the probe gently to create a homogeneous sa...

Страница 18: ...from 848 S to 1978 S 2 If the temperature t C of the conductivity calibration solution is below 0 C or above 50 C 0 C t C 50 C an error message Err 2 will be displayed when performing the auto calibra...

Страница 19: ...eyond the upper steel band see Figure 6 Stir the probe gently to create a homogeneous sample Allow time for the reading to stabilize 4 Press to enter conductivity calibration mode The CA indicator wil...

Страница 20: ...temperature measurement mode See the Change Conductivity toTemperature Measurement Mode section 3 Press to start the temperature calibration process 4 Dip the probe into a solution with known temperat...

Страница 21: ...with automatic or manual temperature compensation With Automatic Temperature Compensation ATC For ATC make sure the phono jack of the probe see Figure 7 is securely inserted The conductivity reading d...

Страница 22: ...k for temperature measurement is disconnected from the meter See Figure 7 2 Switch on the meter and if necessary press to select temperature measurement mode See the Change Conductivity toTemperature...

Страница 23: ...y To avoid contamination or dilution of your sample rinse probe with a small volume of your sample liquid 2 Press to switch on meter 3 Dip the probe into the sample 4 Allow time for the reading to sta...

Страница 24: ...ment range The five ranges are Conductivity Range 0 20 00 S cm 0 200 0 S cm 0 2000 S cm 0 20 00 mS cm 0 200 0 mS cm Note If the value of the solution you are measuring is higher than the range selecte...

Страница 25: ...be used any time in measurement mode 1 To hold a measurement press while in measurement mode HO will appear on the display 2 To release the held value press again Continue to take measurements Note T...

Страница 26: ...urned off 2 Press and simultaneously holding both keys for 2 seconds First release and then release 3 StUP indicator will appear momentarily and CELC will appear next 4 Overview of the setup menu Ente...

Страница 27: ...Select normalization temperature Choice of either 20 C or 25 C Default value is 25 C Select single point calibration Choice of YES or no Default value is YES User re set to factory defaults Choice of...

Страница 28: ...range measurements below 20 S The cell included with your meter has a cell constant of K 1 0 1 Enter the advanced setup as described in the Advanced Setup Overview section 2 Press or until CELC appear...

Страница 29: ...e 1 Selection of Automatic or Manual Calibration In the manual calibration mode you can use your own customized conductivity calibration standard to calibrate this meter 1 Enter the advanced setup as...

Страница 30: ...t of your solution lets you accurately compensate temperature for almost any solution You can adjust 0 0 to 3 0 per C Meter default is 2 1 per C 1 Enter the advanced setup as described in the Advanced...

Страница 31: ...erature of either 25 C or 20 C The default value is 25 C 1 Enter the advanced setup as described in the Advanced Setup Overview section 2 Press or until t nr C appears on the LCD Press 3 Press or to s...

Страница 32: ...es By selecting no to single point calibration you can perform calibration for each conductivity range 1 Enter the advanced setup as described in the Advanced Setup Overview section 2 Press or until S...

Страница 33: ...anged Important Once activated the meter s settings and calibration data will be erased Always exercise caution as meter reset is not reversible 1 Enter the advanced setup as described in the Advanced...

Страница 34: ...not immerse the probe in oily solutions Clean the electrode thoroughly by stirring it in a mild detergent bath or isopropyl alcohol Wipe the probe with soft tissue paper Rinse thoroughly in tap water...

Страница 35: ...rt batteries with correct polarity c Weak batteries c Replace batteries Unstable readings a Air bubbles in probe a Tap probe to remove bubbles b Dirty probe b Clean the probe and recalibrate c Probe n...

Страница 36: ...rr 2 Temperature calibration error Auto calibration is performed outside the temperature range 0 50 C Check the temperature and make sure that it is within the acceptable range If message persists con...

Страница 37: ...nt 0 1 1 0 10 0 selectable Temperature Compensation Automatic Manual from 0 to 50 C Temperature Coefficient 0 0 to 3 0 C Normalization Temperature 20 0 C and 25 0 C selectable Number of Calibration Po...

Страница 38: ...stic body with ATC and BNC connector K 1 0 cm 1 1 meter cable length Calibration Solutions Cat No Description 011008 100 S cm conductivity TDS standard 5 x 60 mL bottles 01100910 147 S cm conductivity...

Страница 39: ...a liquid will conduct more current Conductance derives from Ohms law E IR and is defined as the reciprocal of the electrical resistance of a solution C 1 R where C is conductance siemens R is resistan...

Страница 40: ...are reported as Siemens cm since the value is measured between opposite faces of a cell of a known cubic configuration With most aqueous solutions conductivity quantities are most frequently measured...

Страница 41: ...e per degree centigrade This temperature coefficient change per degree depends on the composition of the solution being measured However for most medium range salt concentration in water 2 per degree...

Страница 42: ...system contact Technical Support Within the United States call 1 800 225 1480 outside the United States call 978 232 6000 or fax 978 232 6031 In Europe the Middle East and Africa contact your local au...

Страница 43: ...portable meter Directive Standard 89 336 EEC Electromagnetic Compatibility EMC Directive EMC essential requirements for measurement device EN 61326 1997 EN 55022 EN 61000 4 2 3 These products have be...

Страница 44: ...following symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further i...

Страница 45: ...once a week to get specified 1 F S accuracy If you are measuring in the mid ranges and you washed the probe in deionized water and stored it dry calibrate the meter at least once a month Wet the probe...

Страница 46: ...ensation by following instructions as described in the Temperature Coefficient section 2 Wait for 5 minutes Note T1 and CT1 conductivity at T1 3 Condition the sample solution and probe to a temperatur...

Страница 47: ...ific Environmental Instruments Water Analysis Instruments 166 Cummings Center Beverly MA 01915 USA Tel 978 232 6000 Toll Free 800 225 1480 Dom Fax 978 232 6015 Int l Fax 978 232 6031 www thermo com wa...

Отзывы: