Thermo Scientific Russell RL060C Conductivity Meter User Guide
1
Without ATC (Manual Temperature Compensation)
For .manual .temperature .compensation, .simply .unplug .the .probe’s .phono .jack .(not .
BNC) .from .the .meter . .
To .use .manual .temperature .compensation, .you .need .to .enter .the .temperature .value .
of .your .process .into .the .meter . .This .is .the .value .at .which .the .reading .will .manually .
temperature .compensates . .You .can .select .any .temperature .between .0 .and .50 .°C . .
(32 .to .122 .°F) . .Default .value .is .25 .°C . .
1 . . .Make .sure .that .the .phono .jack .(for .temperature .
measurement) .is .disconnected .from .the .meter . . .
See .
Figure 7
. .
2 . . .Switch .on .the .meter .and .if .necessary, .press .
.
to .select .temperature .measurement .mode . .See .
the .
Change Conductivity to Temperature
Measurement Mode
section . .
3 . . .Press .
.to .start .temperature .calibration .process . .
4 . . .The .“
CA
” .will .appear .momentarily .and .a .temperature .
value .will .start .flashing . .
5 . . .Check .the .temperature .of .your .sample .using .an .accurate .
thermometer . .Wait .for .the .value .to .stabilize .and .press .
.or .
.and .adjust .the .value .to .the .reference .
thermometer .used . .
6 . . .Press
. .The .[
CO
] .indicator .will .appear .for .1 .5 .seconds, .
and .the .reading .will .stop .flashing . .The .temperature .
calibration .is .successfully .performed . .The .meter .returns .
to .measurement .mode . .