Thermo Scientific Russell RL060C Conductivity Meter User Guide
Manual Calibration
In .manual .calibration .mode, .you .can .use .customized .conductivity .calibration .standards .
(specific .to .your .own .application) .and .calibrate .the .meter . .The .following .example .
shows .calibration .sequence .to .12 .00 .mS .conductivity .calibration .standard . .
1 . . If .necessary, .press .
.to .select .conductivity .mode . .
2 . . .Rinse .the .probe .thoroughly .with .deionized .water .or .
a .rinse .solution, .then .rinse .with .a .small .amount .of .
calibration .standard .
3 . . .Dip .the .probe .into .the .calibration .standard . .Immerse .the .
probe .tip .beyond .the .upper .steel .band .(
see Figure 6
) . .
Stir .the .probe .gently .to .create .a .homogeneous .sample . .
Allow .time .for .the .reading .to .stabilize .
4 . . .Press .
.to .enter .conductivity .calibration .mode . .The .
[
CA
] .indicator .will .appear .for .1 .5 .seconds, .and .a .value .will .
appear .flashing .
Note:
To exit calibration without confirmation, press
.
.
again to go back to measurement mode.
5 . . .Wait .for .the .value .to .stabilize .(the .display .will .continue .
flashing) .and .press .
.or .
.and .adjust .the .value .to .the .
calibration .standard .used . .
6 . . .Press .
. .The .[
CO
] .indicator .will .appear .for .1 .5 .seconds, .
and .the .calibration .is .successfully .performed . .The .meter .
returns .to .measurement .mode . .
7 . . .To .perform .the .next .point .calibration .in .the .multi-point .
calibration .for .next .range, .repeat .step .1-6 .again .until .all .
points .have .been .calibrated .