Thermo Scientific Russell RL060C Conductivity Meter User Guide
26
Temperature Coefficient
The .temperature .coefficient .is .the .amount .of .change .in .conductivity .per .degree .
of .temperature; .it .is .expressed .in .percent .per .°C . .Entering .the .exact .temperature .
coefficient .of .your .solution .lets .you .accurately .compensate .temperature .for .almost .any .
solution . .You .can .adjust .0 .0 .to .3 .0 .% .per .°C . .
Meter .default .is .2 .1% .per .°C .
1 . . .Enter .the .advanced .setup .as .described .in .the .
Advanced Setup Overview
section . .
2 . . .Press .
.or .
.until .
t.Co %
.appears .on .the .LCD . .
Press .
. .
3 . . .Press .
.or .
.to .select .a .value .between .0 .0 . .
and .3 .0 . .
4 . . .Press .
.to .select . .The .meter .will .take .you .back .to .
the .menu, .
t.Co %
. .
5 . . .Pressor .
.or .
.to .move .to .the .next .menu .or .
press .
.to .exit .to .measurement .mode .
%
%
%