Thermo Scientific Russell RL060C Conductivity Meter User Guide
Automatic Calibration
In .the .automatic .calibration .mode, .the .meter .is .capable .of .accepting .one .to .four .points .
for .multi-point .calibration .with .maximum .of .one .point .per .specific .measurement .range . .
For .the .known .calibration .standard .values .refer .to .
Table 1
. .
1 . . .If .necessary, .press .
.to .select .conductivity .mode . .
2 . . .Rinse .the .probe .thoroughly .with .deionized .water .or .
a .rinse .solution, .then .rinse .with .a .small .amount .of .
calibration .standard .
Note:
For automatic calibration you must use one of the
calibration standards listed in
Table 1
.
3 . . .Dip .the .probe .into .the .calibration .standard . .Immerse .the .
probe .tip .beyond .the .upper .steel .band .(
see Figure 6
) . .
Stir .the .probe .gently .to .create .a .homogeneous .sample . .
Allow .time .for .the .reading .to .stabilize .
4 . . .Press .
.to .enter .conductivity .calibration .mode . .The .
[
CA
] .indicator .will .appear .for .1 .5 .seconds, .and .a .value .will .
appear .flashing .
Note:
To exit calibration without confirmation, press
.
again to go back to measurement mode.
5 . . .Wait .for .the .value .to .stabilize .(the .display .will .continue .
flashing) .and .press
. .The .calibration .standard .value .
will .appear .for .3 .seconds . .If .the .calibration .is .successfully .
performed, .
donE
.will .be .displayed .for .about .3 .seconds, .
and .the .meter .will .return .to .the .measurement .mode . .
mS
mS
mS
mS