17
Fit an oil mist filter or pipe the exhaust gases of the vacuum pump to a chemical
fume hood.
All Thermo
Electron
high vacuum pumps are supplied with a gas ballast oil return kit
(GBORK) and EMF10 oil mist filter. Refer to specific manuals received with your
Thermo
Electron
pump to install these items.
3.6
PUMP OUTLET SAFETY
WARNING: Never engage the vacuum pump when water is present in the Ice
Condenser. The introduction of water directly into the internal mechanism of
the vacuum pump will cause significant damage to the pump.
CAUTION:
Do not attempt to use the ModulyoD if it fails the installation test.
If you do, poor performance may result in the loss of the sample being
freeze dried.
Use the following procedure to test the ModulyoD:
If any of the checks below fail, turn the mains switch (item 4, Figure 2) off, then switch
off the external electrical supply and disconnect the supply from the ModulyoD.
Contact your supplier for advice. Do not attempt to use the ModulyoD.
3.7
TEST AFTER INSTALLATION
2
1
3
4
9
10
5
6
7
8
1. Vacuum pump connector
2. Elbow
3. Steel tube
4. Vacuum hose
5. Nozzle connector (optional)
6. Centering-ring and inlet filter assembly
7. Clamp (optional)
8. Thermo
Electron
LyoPump
TM
9. Alternative elbow position
10. Alternative steel tube position
Figure 9 — CONNECT A VLP195FD OR
VLP120 PUMP TO THE M
ODULYO
D