10
Figure 1 (Page 4) and Table 2 locate and describe, in detail, the function of the drain out-
let, drain valve and vacuum port. Figure 2 and Table 2 locate and describe the electrical
connections.
1.2.1.5 CONNECTIONS
The ModulyoD is shipped with the condenser extension assembly installed (see Figure 8).
This extension assembly is necessary to prevent the formation of ice in the vacuum pump
connection. The extension assembly should only be removed when working with the Spin
Freezer 96 freeze drying accessory (part numbers F05637000 or F05637115).
1.2.1.6 CONDENSER EXTENSION ASSEMBLY
Table 2 –
DRAIN OUTLET DESCRIPTION
Connection
Figure #
Item #
Use
Vacuum port
1
7
To connect your vacuum pump to the ModulyoD
Drain valve
1
6
Close the valve (turn it fully clockwise) to isolate the
condenser chamber from the drain outlet. Open the
valve (turn it anti-clockwise) to admit air into the
vacuum system, and to drain water from the condenser
chamber after it has been defrosted.
Drain outlet
1
5
This outlet is used to drain water from the ModulyoD
during defrosting of the condenser chamber. The outlet
is also used to admit air into the vacuum system.
Power IEC 320 inlet
2
1
To connect the ModulyoD to an external power source.
Auxiliary power
2
2
To connect your vacuum pump to the IEC 320 outlet
power supply which will enable the pump to be
operated directly from the ModulyoD.
RS232 port
2
3
To connect the ModulyoD to a computer or chart
recorder.