16
WARNING:
Ensure that the electrical installation of the ModulyoD conforms
with your local and national safety requirements. It must be connected to a
suitably fused and protected electrical supply and a suitable earth (ground)
point.
1.
Make sure that the ModulyoD is suitable for use with your electrical
supply voltage and frequency.
2.
Locate the standard 2 meter long power cord IEC 320 plug and
outlet supplied with the ModulyoD.
3. Connect a 15 A fuse/breaker at the electrical supply outlet to
protect the ModulyoD
.
3.3
CONNECT THE M
ODULYO
D TO THE ELECTRICAL SUPPLY
WARNING:
If intending to freeze dry product containing sodium azide, make
sure that the vacuum pump and pipeline are suitable for freeze drying these
products. If they are not suitable, there is a severe risk of explosion.
CAUTION:
A two-stage vacuum pump must be used with the ModulyoD. If
not, the pump will have a very short working life.
Connect the vacuum pump to the vacuum pump connector on the front of the
ModulyoD (Figure 1, item 7). The connector is suitable for use with a 10 mm inside
diameter pipeline.
We recommend the use of a Thermo
Elec
t
ron
LyoPump
TM
high vacuum pump with the
ModulyoD. These pumps are suitable for freeze drying products which contain sodi-
um azides.
If you do not use a Thermo
Electron
vacuum pump, the pump must meet the require-
ments specified in Section 2.5. To connect another type of pump to the ModulyoD,
refer to the instruction manual supplied with the pump.
3.4
CONNECT THE VACUUM PUMP
If you have a Thermo
Elec
t
ron
high vacuum pump, use the following procedure to
connect the pump to the ModulyoD. Otherwise, you may order the additional con-
nection kit, which consists of NW25 nozzle connector (item 5, Figure 9) and NW25
clamp (item 7, Figure 9). After having the additional kit installed on your pump, fol-
low the same procedure below.
1.
Remove the vacuum pump connection kit, which is already pre-
assembled, from its packing material.
2.
Push the elbow (item 2, Figure 9) of the connection kit onto the
vacuum port of the ModulyoD (item 1, Figure 9).
3.
Push the other end of the connection kit onto the nozzle connector
of the vacuum pump. It is recommended that you use the ModulyoD
(which has an IEC 320 outlet on the rear part of the unit (item 2,
Figure 2) as a power supply for our vacuum pump, using the power
cord supplied with the ModulyoD. This will enable operation of your
vacuum pump directly from the ModulyoD control panel.
3.5
CONNECT A VACUUM PUMP TO THE M
ODULYO
D