Chapter 8
Hazardous Area Installation
This chapter covers requirements for installing the instrument in hazardous
area applications. Hazardous area certification for North America is
provided by the Canadian Standards Association (CSA). Certification for
Europe is provided by Laboratoire Central des Industries Electrique
(LCIE).
General
The area classification determines the type of enclosure to be used and if
barriers are required. If barriers are required, they are installed by the
factory to meet the appropriate agency certification. The standard enclosure
for the instrument is a NEMA 4X (IP65). The instrument is also available
in an optional, factory-installed NEMA 7 explosion proof enclosure. If
transducers are approved for use in hazardous areas, a log of the certifying
agency is affixed to the transducers.
North American
Certification
The transducers are certified by CSA as follows:
●
Intrinsically safe (IS) transducer for Class I and II, Div. 1, Groups
C, D, E, F, G (IS barriers required)
●
Non-incendive for the following areas (IS barriers not required):
- Class I, Div. 2, Groups A, B, C, D
- Class II, Div. 2, Groups E, F, G
The table below lists CSA’s hazardous area installation requirements for
North America.
Table
8–1.
North American hazardous area installation requirements
Item
Division 1
Division 2
Unclassified
Instrument
NEMA 7 enclosure
1
NEMA 4X enclosure
NEMA 4X enclosure
Transducers
IS, barriers required
Non-incendive,
barriers not required
Barriers not required
Installation drawing
2
Figure 8–1
Figure 8–2
N/A
1
Instrument should be installed in accordance with National Electrical Code (NEC) Article 500, including use
of explosion proof seals for the wiring connections to the enclosure.
2
The flowmeter must be installed and wired in accordance with the specified installation drawing.
Thermo Fisher Scientific
DCT6088 User Guide
8-1
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...