System Installation
Transducer Mounting Methods
14.
Remove the paper from the pipe and trace the shape of the 9 o’clock
transducer in the same manner you did for the 3 o’clock transducer.
Ensure that the inside edge of the transducer (opposite the cable
connection) is even with the point just marked on the 9o’clock side of
the pipe.
Figure
2–18.
Z mounting method, step 14
15.
Mount the transducers with pipe straps by following steps 5-6 in
“Spacing & Mounting the Transducers”
earlier in this chapter.
Figure
2–19.
Z mounting method, step 15
The figure below illustrates the final Z method installation.
Figure
2–20.
WV and WW
Methods
For applications with pipe diameters smaller than 2 inches (50 mm), the
WV and WW methods are options that allow higher accuracy and stability
to be achieved when reasonable signal strength can be obtained (10% or
higher).
2-12
DCT6088 User Guide
Thermo Fisher Scientific
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...
Страница 118: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...