Primary Displays & Menus
The 4 –20 mA Menu
3.
Press
Enter
, and a prompt to enter the month will be displayed. Press
Enter
after making the entry. Prompts to enter the day, year hour,
minute, and second will be displayed. Press
Enter
after each entry.
Figure
5–45.
4.
To keep the current values displayed in any screen, scroll to the next
screen with the
down arrow
instead of pressing
Enter
.
5.
Once all parameters are entered, the new programmed date and time
will be displayed.
The 4 –20 mA
Menu
Under the 4–20 mA submenu is the Current Loop Group menu (56),
which contains three setup menus: Span, Cal., and Test (menus 57 through
59).
Figure
5–46.
Current Loop Group menu
The 4–20 mA current loop is factory calibrated and should not require
field calibration prior to use. If calibration and testing should become
necessary, complete the following procedure that follows.
1.
Connect a milliammeter to the input (IN) and output (OUT)
terminals of the current loop module.
2.
Access the Current Loop Calibration menu (58).
Figure
5–47.
4 mA Calibrate screen
5-18
DCT6088 User Guide
Thermo Fisher Scientific
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...
Страница 118: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...