Chapter 3
Wiring
Cable Routing
able Routing
Warning
Avoid potential injury or damage to equipment. Ensure the
flowmeter is grounded at ALL times.
▲
Warning
Minimize the possibility of explosion in hazardous areas. Do not
disconnect power until the area is known to be non-hazardous.
▲
Warning
Prevent the possibility of electrical shock or damage to the
instrument. Disconnect power prior to removing the sheet metal access
cover. Replace the cover before reconnecting power to the unit.
▲
Warning
All wiring should be routed through conduit or cable glands to
seal the enclosure. Refer to Figure 3–1 for the recommended cable routing.
Bonding between conduit connections is not automatic and must be
provided as part of the installation.
▲
Thermo Fisher Scientific
DCT6088 User Guide
3-1