Chapter 7
Maintenance & Troubleshooting
Replacing the
Fuse
placing the
Fuse
Warning
Prevent possible electrical shock and / or damage to the meter.
Disconnect power to the meter prior to performing this procedure.
▲
The flowmeter has a single fuse that is mounted to the main board. The
sheet metal access cover does not need to be removed since there is a fuse
access hole. To replace the fuse:
1.
Disconnect power from the flowmeter.
2.
Determine the cause of the fuse failure and correct if known.
3.
Open the door of the flowmeter.
4.
Using a small, flat-bladed screwdriver, remove the fuse and replace it
with another fuse of the same rating.
5.
Reconnect power to the flowmeter and verify that the unit is operating
properly, e.g. the replacement fuse does not blow.
Thermo Fisher Scientific
DCT6088 User Guide
7-1
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...
Страница 118: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...