Primary Displays & Menus
The Calibration Menu
The best method of zeroing the instrument is to stop the flow and perform
a zero flow set on the pipe. The purpose of the zero flow set is to zero the
instrument for the individual application. This method is used only when
flow in the pipe can be stopped. The flow rate displayed in the
Flow/Velocity primary display (01) must be between -0.25 and+0.25 ft/s
(-0.076 and +0.076 m/s).
Zero Flow Set Method
1.
Ensure there is no flow in the pipe.
2.
Access the Zero Set menu (51).
Figure
5–38.
3.
Select
No Flow
. If the zero flow set calibration is successful, Figure
5–39 is displayed. If the flow rate is too high to perform a zero flow set
calibration, Figure 5–40 is displayed. In this case, continue to step 4.
Figure
5–39.
Zero set calibration successful
Figure
5–40.
Zero set calibration failed
4.
If Figure 5–40 is displayed, press any key and repeat steps 1 through 3
when the flow rate is within required limits.
5-14
DCT6088 User Guide
Thermo Fisher Scientific
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...
Страница 118: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...