System Installation
Small Pipe Applications
2-14
DCT6088 User Guide
Thermo Fisher Scientific
Table
2–5.
Pipe Sizes
Spacing
3.0 to 3.5 inches (76.2 to 88.9 mm)
0.50 inches (12.7 mm)
2.5 to 3.0 inches (63.5 to 76.2 mm)
0.44 inches (11.2 mm)
2.0 to 2.5 inches (50.8 to 63.5 mm)
0.38 inches (9.7 mm)
1.5 to 2.0 inches (38.1 to 50.8 mm)
0.32 inches (8.13 mm)
1.5 inches (38.1 mm ) and smaller
0.25 inch (6.35 mm)
Figure
2–22.
3.
Remove the adhesive protection paper to expose the tape strips. Place a
strip on each side of the surface along the line. Press the strips down to
ensure good adhesion.
Figure
2–23.
4.
Apply coupling compound to the space between the tape strips. The
optimum height of the compound layer is approximately half the
height of the tape strips.
Figure
2–24.
Содержание DCT6088
Страница 2: ......
Страница 3: ...DCT6088 Dedicated Transit Time Flowmeter User Guide P N 1 0561 006 Revision J ...
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 12: ...This page intentionally left blank ...
Страница 20: ...This page intentionally left blank ...
Страница 78: ...This page intentionally left blank ...
Страница 94: ...This page intentionally left blank ...
Страница 97: ...Pipe Schedules Table A 2 Cast iron pipe standard classes Thermo Fisher Scientific DCT6088 User Guide A 3 ...
Страница 98: ...Pipe Schedules A 4 DCT6088 User Guide Thermo Fisher Scientific Table A 3 Ductile iron pipe standard classes ...
Страница 118: ...This page intentionally left blank ...
Страница 120: ...This page intentionally left blank ...