Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of Concentration
34
C. Model 720Aplus Concentration Calibration
and Measurement
Concentration Calibration Procedure
1. The number of the electrode input in use will appear on the prompt
line. Example: CH-1. To select a different channel press 2nd
then channel.
2. Press mode until the concentration mode indicator CON appears on
the display.
3. Add ionic strength adjuster, or pH adjuster to the standard and
samples,as recommended in the electrode instruction manual.
NOTE: If you want to exit this menu, at any time, press
1st then measure.
4. Press calibrate. CALIBRATE will be displayed as well as the time
and date of the last calibration.
5. When ENTER NO. STDS prompt appears enter the number of
standards to be used in the calibration and press yes.
6. Rinse electrode(s) and place into first standard (the least concentrated
standard). Wait for stable display.
7. When the READY ENTER VALUE prompt appears, enter the
concentration value of the standard and press yes.
8. If two or more standards were selected, repeat steps 6 and 7 for
each standard.
9. If one standard was selected, the electrode slope in mV/decade will be
displayed at the SLOPE prompt. Press yes if correct, or enter the
correct value for the electrode.
NOTE: Make sure the correct polarity sign is entered
for the electrode in use during a one point calibration.
The sign should be + (positive) for cations, for example,
Na
+
, and - (negative) for anions, for example F
-
.
10. If two or more standards were entered, the electrode slope is
calculated and displayed.
11. The meter then advances to MEASURE mode.
12. Rinse and place electrode(s) into the sample. Record concentration
directly from the meter display when the READY prompt appears.
Temperature is displayed in the prompt line.
Содержание 410Aplus
Страница 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Страница 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Страница 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Страница 115: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Страница 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Страница 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...