G. Models 520Aplus, 525Aplus, 720Aplus Dissolved
Oxygen Measurements
Dissolved Oxygen measurements are displayed in ppm when the Thermo
Orion 9708 Dissolved Oxygen Electrode (Cat. No. 970899) is used with
the Thermo Orion Models 520Aplus, 525Aplus and 720Aplus and
920Aplus Meters. Perform the following steps to prepare the meter and
calibrate the electrode.
1. Connect the D.O. electrode Model 9708 to meter. Leave electrode
mode switch “OFF”.
2. For the Models 525Aplus and 720Aplus select the correct channel by
pressing 2nd then channel.
3. Disconnect ATC probe.
NOTE: ATC probe must not be connected to the meter
and the meter should be calibrated to pH 7 at 25.0˚C.
4. Press mode until pH mode indicator is displayed.
5. Press hold. Turn the hold feature off.
6. Press measure. Press set temp. Enter 25.0 at the prompt and
press yes.
7. Press calibrate. Enter 1 at the ENTER NO. OF BUFFER prompt
and press yes.
8. Enter 7.00 to the CAL AS prompt and press yes.
9. At the slope prompt enter 100.0 and press yes. The meter
automatically returns to measure mode.
10. Turn the mode switch on the electrode to BT CK. Good battery
operation is indicated by a reading of 13.40 or greater on the meter.
11. Turn the mode switch on the electrode to ZERO. Use the zero
calibration control on the electrode to set the meter to read 0.00.
Benchtop pH/ISE Meter Instruction Manual
Other Measurement Modes
54
Содержание 410Aplus
Страница 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Страница 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Страница 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Страница 115: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Страница 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Страница 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...