Benchtop pH/ISE Meter Instruction Manual
Overview of Meter Functions 520Aplus,525Aplus,720Aplus and 920Aplus 109
Automatic Datalog Function
The meters have an internal datalogger that will record and store a
maximum of 25 points automatically. The data points remain in memory
until cleared.
Prompt
Keypress/Action
Press data/print modes
1. MANUAL PRINT
The current print mode is displayed. Wait
for the prompt.
2. SET PRINT MODE ?
Press no.
3. DATALOG
The current datalog setting is displayed.
MANUAL
Wait for prompt.
4. SET LOG MODE
Press yes.
5. 0-DISABLE 1-RDY
The datalog setting options are presented
2-TIM 3-MAN
two at a time. Press no to display the other
two. Enter 0, 1, 2 or 3, then press yes.
0 To disable datalogging.
1 For datalogging on ready.
2 For datalogging at present time
interval.
3 For manual datalogging when print
is pressed.
6. DATALOG
The meter verifies the setting and returns to
MANUAL
measure mode.
7. TIMED LOG
If 2 (datalog at a preset time interval) is
selected the interval must be entered at the
following prompts.
8. DATALOG INT HR
Enter the datalog interval hours setting, from
0 to 23, then press yes.
9. DATALOG INT MIN
Enter the datalog interval minutes setting,
from 0 to 59, then press yes.
10. DATALOG INT SEC
Enter the datalog interval seconds setting
from 0 to 59, then press yes. Logging can
not take place faster than every 5 seconds.
11. DATALOG
The datalog interval is displayed then the
INTERVAL
meter returns to MEASURE mode.
Содержание 410Aplus
Страница 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Страница 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Страница 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Страница 115: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Страница 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Страница 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...