![Thermo orion 410Aplus Скачать руководство пользователя страница 118](http://html1.mh-extra.com/html/thermo-orion/410aplus/410aplus_instruction-manual_1104441118.webp)
Benchtop pH/ISE Meter Instruction Manual
Multipoint Calibration Curve 117
Appendix E
Multipoint Calibration Curve
Models 410Aplus, 420Aplus, 520Aplus, 525Aplus, 710Aplus, 720Aplus,
and 920Aplus utilize a “point-to-point” scheme for multipoint calibrations
in both pH and CONCENTRATION modes. In this technique, straight line
segments are “drawn” between each successive pair of calibration points.
Separate slope and E
o
values are calculated using the adjacent pair of data
points. For example, in a five point calibration, four slopes and five E
o’
s are
calculated to correspond to the four line segments that connect the data.
Concentration or pH values for samples are calculated according to which
segment their potential values fall into. Those above or below the range of
calibration are calculated by extrapolation from the top or bottom most
segment.
The displayed slope is an average slope of the separate segments. The slope
may be outside the range normally expected for an electrode. For example,
if the slope of one segment is low, then the average will be lower than
theoretical, even though the response may be Nernstian over most of
the range.
See Figure below for an example of a typical calibration curves using
this method.
250
200
150
100
50
0
-50
-100
-150
-200
-250
mV
pH3 pH4 pH5 pH6 pH7
pH8 pH9 pH10
Segmented Calibration (
)
and Best Fit Line ( )
Содержание 410Aplus
Страница 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Страница 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Страница 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Страница 115: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Страница 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Страница 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...