Benchtop pH/ISE Meter Instruction Manual
Calibration and Measurement of Concentration
31
Chapter VI
Calibration and Measurement of
Concentration
Models 710Aplus, 720Aplus, 920Aplus
A. General Information
A one, two, or multipoint calibration should be performed before
concentration is measured. It is recommended that a two point standard
calibration be performed at the beginning of each day and every time
electrodes are changed to determine the slope of the electrode. This serves a
dual purpose, determining if the electrode is working properly and storing
the slope value into memory. Perform a calibration with one standard every
two hours to compensate for possible electrode drift. Use a fresh aliquot
from one of the standards used in the initial calibration. During calibration,
always use the most dilute standard first, the meter will automatically
recognize slope direction (i.e. will recognize anion or cation electrodes).
Standards should bracket the sample range and be in the same
concentration units.
Units
Any convenient units of concentration units can be used. For example:
molarity, ppm, %, etc.
Temperature
Allow all samples and standards to reach the same temperature before
measurement, or use temperature compensation if the electrode isopotential
point is known. (Model 920Aplus only)
Содержание 410Aplus
Страница 3: ...2 Benchtop pH ISE Meter Instruction Manual...
Страница 81: ...80 Benchtop pH ISE Meter Instruction Manual Certificate of Conformity...
Страница 89: ...Benchtop pH ISE Meter Instruction Manual Ordering Information 88...
Страница 101: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 420Aplus 710Aplus 100...
Страница 115: ...Benchtop pH ISE Meter Instruction Manual Overview of Meter Functions 520Aplus 525Aplus 720Aplus and 920Aplus 114...
Страница 117: ...Benchtop pH ISE Meter Instruction Manual Meter Default 116...
Страница 119: ...Benchtop pH ISE Meter Instruction Manual Multipoint Calibration Curve 118...
Страница 123: ...Benchtop pH ISE Meter Instruction Manual Automatic Blank Correction 122...