72
2-6-5. Input / Output Trigger Settings
2-6-5-1. Trigger in Delay
Description
The Trig in Delay setting determines how long to delay
any action after a trigger is received.
Default setting 0.0us
Operation
Press
Shift
>
Utility
Help
>
Other
[F5] in order,
and set the Trig in Delay setting.
Trig in Delay:
0.0
– 5000us
2-6-5-2. Trigger Out Width
Description
The Trigger Out Width setting sets the trigger output
signal’s pulse width.
Default setting 10.0us
Operation
Press
Shift
>
Utility
Help
>
Other
[F5] in order,
and set the Trig Out width.
Trig Out width:
2.5
– 5000us
2-6-6. Measure Average
Description
The Measure Average setting is used to set the speed of
the measurement display. The setting has three modes.
They are slow, normal and fast
The default mode for Measure Average setting is slow.
Operation
1. Press
Shift
+
Utility
Help
>
Other[F5]
.
2. Set the
Measure Average
setting.
Slow
Normal
Fast
Default
Average 64 times; Display spend time:1280ms
Average 16 times; Display spend time:320ms
Average 4 times; Display spend time:320ms
Slow mode
Содержание LSG-1050H
Страница 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Страница 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Страница 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Страница 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Страница 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Страница 204: ......