116
Power Source
Config Error
Config Error indicates that the power supply voltage is
less than the Trig V voltage setting after the test has
started. This can indicate that the power supply output is
not on or that the power supply output or Trig V is
incorrectly configured.
FSEQ
RS232
03/Oct/2018
OCP Test
0.00
V
0.000
A
0.00
w
Config Error
Config Error indicates that upon
starting the power supply voltage is
less than the Trig V voltage.
Set Curr:
Fetch Volt:
Note
In addition to the OCP settings as described above, the
Trig Voltage settings must also be set according to the
output characteristics of the power supply.