141
4-1-9. Load On/Off Status
Description
Pins 13 and 17 (Load On Status) of the J1 connector is used
to monitor the load status (on or off).
Pin out
The Load On Status pin is a
photo-coupled open-collector
output.
17
13
Photo-coupler input: 30V max, 8mA, max.
4-1-10. External Control of the Range
Description
The
I Range for the present operating mode can be
externally controlled when the I Range is set to high range.
The range is changed using pins 8, 9 (Range Cont 1 &0) and
12 (A Com) of the J1 connector.
Operation
1.
Press
Main
>
Configure
[F5] >
Next Menu
[F4] >
External
[F3] and set the
Control
setting to
V
,
R
or
Riv
to
enable external control.
2.
When externally controlling the range, the pin input
combination determines which range is chosen.
I Range
Pin 9
Pin 8
H
High
High
M
High
Low
L
Low
High
Pin Inputs
Pins 8 and 9 of the J1 connector are internally pulled up to
5V with a 10kΩ resistor when open. When closed, pin 8 and
9 are pulled down to the A COM ground level.
Connection
LSG
Switches
Analog
connector
12
+5V
A COM
8
9
10k
Ω
Note
The I Range can only be externally controlled when the I
Range has been set to High using the manual operation.
Содержание LSG-1050H
Страница 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Страница 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Страница 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Страница 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Страница 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Страница 204: ......