45
2-1-9. Safety Short
Description
When activated, the safety short function only allows the
short key to be used when the load is already on.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the Short Safety.
When set to OFF, the load can be shorted at any time.
When set to ON, the load can only be shorted when the
load is already on.
Short(Safety):
OFF,ON
Note
The Short Safety setting will be grayed out if Short
Function is set to OFF. See page 45 (
2-1-11.Short Function
Enable/Disable
) for Short Function.
2-1-10. Short Key Configuration
Description
The Short key can be configured to Toggle or Hold. By
Default the Short key is set to Toggle.
Toggle: Pressing the Short key will toggle the shorting
function on or off.
Hold: Holding the short key will short the load.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the
Short Key
setting.
Short Key:
Toggle, Hold
Note
The Short Safety setting will be grayed out if Short
Function is set to OFF. See page 45 (
2-1-11.Short Function
Enable/Disable
) for Short Function.
2-1-11. Short Function Enable/Disable
Description
The short key can be disabled to prevent the operator
accidentally shorting the load.
Operation
Press
Main
>
Configure
[F5] >
Other
[F2] in order,
and set the
Short Function
.
When set to OFF, the Short key is disabled and all short
configuration options in the Main > Configure> Other
menu are also disabled.
When set to ON, the Short key is enabled.
Short Function:
OFF,ON
Содержание LSG-1050H
Страница 24: ...14 GRA 414 E EIA standard 132 382 5 477 6 GRA 414 J JIS standard 149 382 5 477 6...
Страница 185: ...175 7 6 Operating Area 7 6 1 LSG 175H H Range M Range L Range...
Страница 186: ...176 7 6 2 LSG 350H H Range M Range L Range...
Страница 187: ...177 7 6 3 LSG 1050H H Range M Range L Range...
Страница 188: ...178 7 6 4 LSG 2100H H Range M Range L Range...
Страница 204: ......