Control unit
Settings register
39
Setting the alarm limit for differential pressure -80 to 80
mbar
1. Click on differential pressure
-80 to 80 mbar
.
2. Enter the required value in the
“
mbar
” box.
3. The alarm limit is activated by clicking
on the “
Alarm activated
” checkbox.
4. You will get an audible signal by
clicking on the “
Audible signal
”
checkbox.
5. Save the data from the current register
card by clicking on the “
Apply
” button.
6. Click on“
OK
” to save the data and to
close the instrument driver window.
Setting the alarm limit for -1000 to 1000 mbar differential
pressure
1. Click on differential pressure
0...1000 mbar.
2. Enter the required value in the
“
mbar
” box.
3. The alarm limit is activated by clicking
on the “
Alarm activated
” checkbox.
4. You will get an audible signal by
clicking on the “
Audible signal
”
checkbox.
5. Save the data from the current register
card by clicking on the “
Apply
” button.
6. Click on“
OK
” to save the data and to
close the instrument driver window.