©2004 TESA Switzerland All Rights Reserved.
Page 3-8
User's Manual
MH3D
Starting & Stopping
Make sure that the following steps have been taken before starting the machine:
• Check the packing list to ensure that all necessary items have been
received.
• The machine and workstation is in a thermally stable environment (see
Specifications).
• The machine has been leveled such that the bridge does not drift when all
axis lock switches are unlocked and the probe holder is not being held.
• Check to ensure the machine was installed according to the specifications
in the Installation Manual.
• Verify that all shipping brackets have been removed.
• Check the
input air supply
pressure. It should be between 4.8 to 8.3 BAR
(70 and 120 psi) and should provide for a minimum consumption of
21L/min (.74 CFM). The pressure should be relatively constant and not
subject to drastic changes. The air supply should also be relatively free of
contaminants. NOTE: The maximum air pressure is 8.3 Bar (120 psi)
• The air supply is connected and the regulator is factory set to the correct
pressure. The pressure gage displays 3.7 BAR (55±1 psi). The pressure has
been preset at the factory. If the pressure is not correct, unlock the knob at
the bottom of the regulator, adjust the pressure, and relock the knob.
• All cables from the machine to the MH3D controller (ie. encoder, ground,
ZMouse and, if included, TTP, etc.) are connected (reference the
schemtatics).
• Ensure that the power supply and, if included, the printer is properly
connected. For additional information about the printer, refer to the
manufacturer’s technical publications for operating procedures.
• Verify that all machine guards and covers are in place.
• Wipe all exposed ways and the table work surface with alcohol and a
clean, soft cloth to remove dust or residue. (Reference the Cleaning
Bearing Surfaces in Preventive Maintenance Section)
Содержание MH3D
Страница 1: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page 0 1 U S E R S M A N U A L English ...
Страница 2: ... 2004 TESA Switzerland All Rights Reserved Page 0 2 User s Manual MH3D ...
Страница 6: ... 2004 TESA Switzerland All Rights Reserved Page 1 2 User s Manual MH3D ...
Страница 17: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page 2 1 CHAPTER 2 Construction Construction ...
Страница 18: ... 2004 TESA Switzerland All Rights Reserved Page 2 2 User s Manual MH3D ...
Страница 34: ... 2004 TESA Switzerland All Rights Reserved Page 2 18 User s Manual MH3D Cables Power supply Printer ...
Страница 35: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page 3 1 CHAPTER 3 Operation Operation ...
Страница 36: ... 2004 TESA Switzerland All Rights Reserved Page 3 2 User s Manual MH3D ...
Страница 53: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page 3 19 ...
Страница 54: ... 2004 TESA Switzerland All Rights Reserved Page 3 20 User s Manual MH3D Useful Probe Dimensions TESASTAR_i ...
Страница 56: ... 2004 TESA Switzerland All Rights Reserved Page 3 22 User s Manual MH3D MH3D Notes ...
Страница 57: ...Page 4 1 2004 TESA Switzerland All Rights Reserved User s Manual MH3D CHAPTER 4 Maintenance Maintenance ...
Страница 58: ...Page 4 2 2004 TESA Switzerland All Rights Reserved User s Manual MH3D ...
Страница 69: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page 5 1 CHAPTER 5 Glossary Glossary ...
Страница 70: ... 2004 TESA Switzerland All Rights Reserved Page 5 2 User s Manual MH3D ...
Страница 92: ... 2004 TESA Switzerland All Rights Reserved Page 5 24 User s Manual MH3D MH3D Notes ...
Страница 93: ...User s Manual MH3D 2004 TESA Switzerland All Rights Reserved Page A 1 CHAPTER A Appendix Appendix ...
Страница 94: ... 2004 TESA Switzerland All Rights Reserved Page A 2 User s Manual MH3D ...