Guidance and Manufacturer’s Declaration- Emissions
All ME Equipment and ME Systems
Guidance and Manufacturer’s Declaration- Emission
The SmartPReP2 is intended for use in the electromagnetic environment specified below. The customer or user of the SmartPReP2
should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment-Guidance
RF Emissions
CISPR 11
Group 1
The SmartPReP2 uses RF energy only for its internal function. Therefore, its RF emissions are low and are
not likely to cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A or B
A
Harmonics
IEC 61000-3-2
Class A, B, C, D or
N/A
A
Flicker
IEC 61000-3-3
Complies or N/A
Complies
The SmartPReP2 is suitable for use in all establishments,
other than
domestic, and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic
purposes.
Guidance and Manufacturers Declaration – Immunity
All ME Equipment and ME Systems
Guidance and Manufacturer’s Declaration-Immunity
The SmartPreP2 is intended for use in the electromagnetic environment specified below. The customer or user of the SmartPReP2
should ensure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment-
Guidance
ESD
IEC 61000-4-2
±6kV Contact
±8kV Air
±6kV Contact
±8kV Air
Floors should be wood, concrete or ceramic tile. If
floors are synthetic, the r/h should be at least 30%
EFT
IEC 61000-4-4
2kV Mains
±1kV I/Os
2kV Mains
±1kV I/Os
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips/Dropout
IEC 61000-4-11
>95% Dip for
0.5 Cycle
60% Dip for 5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
>95% Dip for
0.5 Cycle
60% Dip for 5 Cycles
30% Dip for
25 Cycles
>95% Dip for
5 Seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the SmartPrep2 requires continued operation during
power mains interruptions, it is recommended that
the SmartPreP2 be powered from an uninterruptible
power supply or battery.
Power Frequency
50/60 Hz
Magnetic Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be that of a
typical commercial or hospital environment.