MAINTENANCE
The SmartPReP 2 System requires no calibration and is designed for minimum maintenance. Internal
software monitors every process run from start to finish. If the timing or RPM vary more than the
specifications listed below the unit will stop and the Interrupted light will illuminate.
Optional - SPEED VERIFICATION
The centrifuge has a set speed protocol with two points that can be confirmed as follows:
Time
Speed should be
A
1-3 minute(s) into process
2500RPM±150 (2350 - 2650)
B
6-9 minutes into process
2300RPM±140 (2160 - 2440)
The following lists the recommended maintenance procedures that should be followed.
CLEANING
EVERY USE
CAUTION: Ensure line cord is disconnected from power source prior to cleaning. Do not use abrasive
cleaning agents. Do not use solvents or strong alcohol solutions. Do not immerse any part of the
SmartPReP 2 System in liquids.
The SmartPReP 2 System should be cleaned after each use and/or after any spill of liquids. For routine
cleaning use only mild detergents, water and a soft damp cloth to thoroughly clean the inside and outside
of the unit. Allow unit to dry before returning to service. A dry cloth or paper towels can be used to dry
the unit.
0
500
1000
1500
2000
2500
3000
1
36 71 106 141 176 211 246 281 316 351 386 421 456 491 526 561 596 631 666 701 736 771 806
R
M
P
Speed Check points
A
B