ELECTRICAL SPECIFICATIONS
Model #
SMP2-100
SMP2-115
SMP2-230
Input Voltage (VAC)
100
115
230
Operating Frequency (Hz)
50/60
50/60
50/60
Rated Current (Amps)
5
5
2.5
Fuse rating
5 Amp, time delay,
glass tube, ¼” x 1¼”
5 Amp, time delay,
glass tube, ¼” x 1¼”
2.5 Amp, time delay,
glass tube, 5 x 20mm
ENVIRONMENTAL CONSIDERATIONS
Temperature limit:
Operational: 10°C - 30°C (50°F - 86°F)
Storage: -40°C - 66°C (-40°F - 150 °F)
Humidity Limit:
Operation: 10-90% non condensing
Storage: 10-90% non condensing
Minimum clearance distance around unit: 3 inches
There is no need to secure the unit with straps or other mounting hardware. Leveling is not required as
long as the unit does not slide off the intended surface.
ELECTROMAGNETIC COMPLIANCE
Special precautions regarding EMC: Do not stack SmartPReP 2 Centrifuges on other medical electrical
equipment. Always check for normal operation when the SmartPReP 2 Centrifuge is moved to a new
location.
Portable and mobile RF communications equipment can effect medical equipment.
WARNING: This equipment/system is intended for use by healthcare professionals only. This
equipment/system may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take mitigation measures, such as reorienting or relocating
the SMP2 or shielding the location.
The use of accessories, transducers and cables other than those supplied by Harvest Technologies, Corp,
may result in increased EMISSIONS or decreased IMMUNITY of the SMP2.