Page
23
of
26
4531200 / 1052
IEC 60601-1-2
Issued: 2007/03/01 Ed: 3.0 Medical Electrical Equipment - Part 1-2: General
Requirements for Basic Safety and Essential Performance - Collateral Standard:
Electromagnetic Compatibility - Requirements and Tests Requirements for Basic
Safety and Essential Performance - Collateral Standard: Electromagnetic
Compatibility - Requirements and Tests Evaluation for Japan requirements.
FCC 47CFR 15B clA
Issued:2011/04/21 Title 47 CFR Part 15 Subpart B Unintentional Radiators Class A
Verification
Harvest Technologies ETL Control #117553
ELECTRICAL SPECIFICATIONS
Model #
SMP2-100
SMP2-115
SMP2-230
Input Voltage (VAC)
100
115
230
Operating Frequency (Hz)
50/60
50/60
50/60
Rated Current (Amps)
5
5
2.5
Fuse rating
5 Amp, time delay,
glass tube, ¼” x 1¼”
5 Amp, time delay,
glass tube, ¼” x 1¼”
2.5 Amp, time delay,
glass tube, 5 x 20mm
ENVIRONMENTAL CONSIDERATIONS
Temperature limit:
Operational: 10°C - 30°C (50°F - 86°F)
Storage: -40°C - 66°C (-40°F - 150 °F)
Humidity Limit:
Operation: 10-90% non condensing
Storage: 10-90% non condensing
Minimum clearance distance around unit: 3 inches
There is no need to secure the unit with straps or other mounting hardware. Leveling is not required as
long as the unit does not slide off the intended surface.
U.S. Patent Nos.: 5,707,331; 5,895,346. US and foreign patents pending. BSI registered firm. ISO 13485
Registered.
ELECTROMAGNETIC COMPLIANCE
Special precautions regarding EMC: Do not stack SmartPReP 2 Centrifuges on other medical electrical
equipment. Always check for normal operation when the SmartPReP 2 Centrifuge is moved to a new
location.
Portable and mobile RF communications equipment can effect medical equipment.