Guidance and manufacturer’s declaration – electromagnetic emissions
(Table 201)
The SMP2 is intended for use in the electromagnetic environment specified below. The customer or the user of the SMP2 should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The SMP2 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SMP2 is suitable for use in all establishments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
(Table 202)
The SMP2 is intended for use in the electromagnetic environment specified below. The customer or the user of the SMP2 should assure that it is used in such an
environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
Lines
± 2 kV for power supply lines
Not applicable for input/output lines,
there are no input/output lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T
)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
>95%-0.5 cycles
60%-5 cycles
30%-25 cycles
>95%-250 cycles
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the SMP2 requires continued operation during
power mains interruptions; it is recommended that the SMP2 be powered
from an uninterruptible power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m @ 50Hz
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE
U
T
is the arc. mains voltage prior to application of the test level.