Page
10
of
26
4531200 / 1052
Loading the Process Disposable – AdiPrep Adipose Transfer System
Note: The Adipose Hardware Kit (AHK-1) is required to operate the SMP2 System with the AdiPrep process disposable.
See Scenarios 1 and 2 for adipose processing.
Install the two adapters into the rotor trunnions
with the white dots aligned as shown.
Be sure the adapter is properly seated
as shown below
Correct Incorrect
To process adipose tissue in the SMP2 System:
Scenario 1 - Install one AdiPrep process disposable
into one AA-1 Adapter and the BW-ADI AdiPrep
Balance Weight into the opposing adapter, as
shown below.
Scenario 2 - Install two (2) AdiPrep process disposables, one in
each AA-1 Adapter, as shown below.
1.
Close the lid.
2.
Press the green START or the blue PRP button.
3.
The TIMING window will display the remaining time of the cycle. When the display indicates 10 (approximately 4
minutes have passed) press the red STOP button.
4.
When the rotor has stopped, press the LID button and remove the PDs.
CONTRAINDICATIONS
The use of the SmartPReP 2 System may be contraindicated when there is: Clinical or laboratory evidence
of septicemia for patients who have taken aspirin, or other medications that alter platelet function, within
3 days prior to surgery, or patients with disorders associated with platelet dysfunction.
WARNINGS
Federal law (USA) restricts this device to sale by, or on the order of a physician. The physician is solely
responsible for the use of this device.
Safe and effective use of this device requires proper set-up and operation by trained personnel.
This device should not be used in the presence of flammable agents.