3F HOT CATHODE MINI-BA MODULE (software ver. 2.31 or higher)
With a hot cathode mini-BA installed in station 5 and when the setup button is pressed, the following appears on
the display,
3F
(AUTO or SELF or BOTH)
Use to up/dn arrow to change between the following,
AUTO
This means CC or BA is turned off by thermoconductivity gauge pressure reading only and comes back on
when pressure falls below 10 microns
If CC or BA exceeds the range, reading should blink until overpressure is removed.
SELF
This means CC or BA is turned off by CC or BA overpressure only. The CC or BA must be reset by the
front panel or by RS232.
BOTH
This means CC or BA is turned off by both the CC or BA overpressure and off and on by the
thermoconductivity gauge pressure reading. If turned off by overpressure, the CC or BA must be reset by
the front panel or by RS232.
Press the setup button again to advance to the following,
FIL
(ON or OFF or RDY)
Use to up/dn arrow to turn the filament on or off.
Press the setup button again to advance to the following,
(OFF or ON) DGAS
Use to up/dn arrow to turn the degas on or off.
Press the setup button again to advance to the following,
(software ver. 2.31 or higher)
(# time) DGAS
If dgas is off then this selection will not appear, if dgas is on, a time in minutes can be selected. Use the
up/dn arrow to select the time. If 0 is selected, the dgas remains on indefinitely.
Press the setup button again to advance to the following,
FIL
NBR (1 or 2)
Use to up/dn arrow to turn the degas on or off.
Press the setup button again to advance to the following,
(# sensitivity) SENS
Use to up/dn arrow to set the sensitivity.
Press the setup button again to advance to the following,
If a hot cathode mini-BA is installed in station 5 and 6, the above setup will repeat for station 6.