502 USE OF PUSHBUTTONS
TEST/OP PUSHBUTTON
RETURN TO NORMAL OPERATION
- A pushbutton labeled TEST/OP is located between the two sets of
UP/DOWN arrows. Pressing this button at the end of any step listed below under SETUP causes the instrument to
return to normal operation. Any settings that have been changed are stored in nonvolatile memory in an EEPROM,
and are the default settings should power be turned off and then back on (See Fig. 4.2).
SELF DIAGNOSTIC TEST
- The TEST/OP pushbutton is used to perform self diagnostics if pressed during normal
operation. When TEST/OP is pressed in normal operation the two station displays go blank. The left-hand large
display now shows "SELF", and the right-hand large display shows "tESt". The displays remain that way until the
self-diagnostic test is complete, which is probably less than a second. During this time all pushbuttons are
disabled.
At the completion of the self-test period, the display starts to flash alternate horizontal and vertical bars. The other
display LEDs turn on with the horizontal bars. This is done so the operator may observe if any display segments
are faulty.