INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:[email protected] web site:www.televac.com
MM-200_im REV M
Page 106 of 160
TELEVAC
DATA OUTPUT
Axxx<cr>
Output each marked station value
every xxx cycles, where xxx is any value
between 1 and 255.
CA<cr>
Cancel the automatic output.
Mx<cr>
Mark station x, where x is any number between 1 and
9, or A for Station
10.
Rx<cr>
One shot reading of station x.
Ux<cr>
Unmark station #x from automatic output.
DISPLAY
DLx<cr>
Station x is displayed on the left side.
DRx<cr>
Station x is displayed on the right side.
FA(cr)
In conjunction with the "FI" command, will inhibit all front panel function,
except station select.
FE<cr>
Enables all front panel functions, unless
inhibited by the rear panel switch.
FI<cr>
Inhibits certain front panel functions.
UM<cr>
Causes Stations to be displayed in millibars.
UP<cr>
Causes stations to be displayed in Pascals.
UT<cr>
Causes stations to display in traditional units.
GAS
GFx<cr>
Clears Station x from gas operation.
GNx<cr>
Sets Station x for gas (Argon) operation.
HOT CATHODE
AH<cr>
Auto Mode For Hot Cathode Gauges.
EH<cr>
Enable Hot Cathode Gauges (shutdown mode).
FF<cr>
3D Filament OFF.
FN<cr>
3D Filament ON.
GF<cr>
3D Degas OFF.
GN<cr>
3D Degas ON.
GNxxx<cr>
3D Degas ON for "xxx" minutes, 255 max.
SWx<cr>
Make sensor switch setting equal "x".
LEAK RATE
ED<cr>
End Leak Rate data stream (but not the test).
EL<cr>
Stops Leak Rate test.
LR<cr>
Starts Leak Rate test. Time = 0.
RL<cr>
Read Leak Rate value periodically. "LR=nnnn"
where nnnn is the leak rate in microns/hr.
OUTPUT FORMATS
R1EL<cr>
0-10V
of 1Es to be logarithmic.
R1En<cr>
0-10V
of 1Es to be linear.