INSTRUMENT MANUAL
VACUUM GAUGE MODEL MM200
160Phone:(215) 947-2500 fax:(215) 947-7464 e-mail:[email protected] web site:www.televac.com
MM-200_im REV M
Page 134 of 160
TELEVAC
800 SPECIFICATIONS
801 CONTROLLER SPECIFICATIONS
» Indoor use
» Attitude up to 2000 m
» Temperature 5° C to 40° C
» Maximum relative humidity 80% for temperatures up tp 31° C decreasing linearly to
50 % relative humidity at 40° C
» Overvoltage category - category ll of lEC 60364-4-443
» Rated pollution degree: 2
» Voltage 115/230 VAC ±10%
» Frequency
50 - 60 Hz.
» Input Current (Max.)
2A @ 115V
1A @ 230V
» Input Power (Max.)
230W
» Calibration medium
Dry Air, (or Nitrogen)
» Oper. temp. (tube)
0 to 65°C
» Setpoint Relays
Assignable to any station 5A 120 VAC or dry circuit
» Output
Analog 0-10 VDC
» Weight (instrument)
5 lbs.
802 CONTROLLER
DIMENSIONS
The controller dimensions and cut out dimensions are given in Figure 8.1
Fig. 8.1 - Dimensions for Modular Gauge