Getting StartedTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
45
Figure 3-10:
Current Loop Option Installed on Motherboard
C
ONVERTING
C
URRENT
L
OOP
A
NALOG
O
UTPUTS TO
S
TANDARD
V
OLTAGE
O
UTPUTS
To convert an output configured for current loop operation to the standard 0 to 5 VDC
output operation:
1.
Turn off power to the analyzer.
2.
If a recording device was connected to the output being modified, disconnect it.
3.
Remove the top cover.
•
Remove the screw located in the top, center of the front panel.
•
Remove the screws fastening the top cover to the unit (both sides).
•
Slide the cover back and lift straight up.
4.
Remove the screw holding the current loop option to the motherboard.
5.
Disconnect the current loop option PCA from the appropriate connector on the
motherboard (see Figure 3-10).
6.
Each connector, J19 and J23, requires two shunts. Place one shunt on the two left-
most pins and the second shunt on the two pins next to it (see Figure 3-10).
7.
Reattach the top case to the analyzer.
•
The analyzer is now ready to have a voltage-sensing, recording device
attached to that output.
8.
Calibrate the analog output as described in Section 5.9.3.2.
Содержание GFC7001T
Страница 100: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 100 ...
Страница 107: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 107 ...
Страница 118: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 118 ...
Страница 136: ...SetupTeledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 136 This page intentionally left blank ...
Страница 195: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 195 ...
Страница 207: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 207 ...
Страница 211: ...Calibration Procedures Teledyne API Model T300 T300M CO Analyzer Teledyne Analytical Instruments 211 ...